|About |Contact | Korean |

 

EChem News  Letter

e-mail address
Company Name
Country

Name


 

100Watt SMARTTM Fuel Cell Test System

 

*         Feature

 

*   Fully Integrated Compact Size

*   Suitable for single cell(PEM,DM)

*   3 Models:
SMART2TM, SMART2PEMTM,SMART2DM
TM

*   Automatic purge gas control

*   External anode & cathode line and cell temperature control

*   Watch-dog function

*   Full automatic by PC control

*   Max 4channels control by one PC

*   Accurate electronic load

*   Various safety functions

*   Powerful software with independent data analysis software         

 

 

설명: Z:\[원아테크]생산제품관련정보\사진-고해상(사용가능)정리되지않은파일은 사진관폴더로\smart\aa\DM3.jpg

100Watt SMART2TM Fuel Cell test system

 

*         Description

The SMART2TM is very compact, however, fully automated and integrated at a attractive price.

Our control and measurement software with powerful graphical user interface makes you easy to operate the SMART2TM.With the exceptional features of the SMART2TM, you can digitize your fuel cell economically and evaluate it easily.

 

*         Standard Configuration List for SMART2TM PEM/DMFC test system

*  Solenoid valve:5ea

*  MFC(2set) for Anode and Cathode

*  Check valve: 6ea

*  3 Way valve: 2ea

*  Membrane type Humidifier: 2ea

*  Automatic water feeding for humidifier: 2ea

*  Pressure sensor: 2ea

*  Temperature controller(with line heater & thermocouples): 7set

*   Humidifier Temperature controller 2set,

*   Instrument inside gas line Temperature controller:2set,

*   Instrument outside gas line Temperature controller: 2set

*   Stack or cell Temperature controller: 1set

*  Temperature monitoring only: 2 points with thermocouples
Anode & cathode gas line inside

*  Liquid Pump for Methanol or Ethanol etc(Piston type) 1set

*  Internal Load bank: 1set

*  Back pressure regulator (Manual type)

*  System controller with emergency switch

*  Control PC(option) with WFTS software

*  Interface boards with cables

 

 

*         Optional Equipment List

*  MFC(s) for Anode and/or Cathode for Reformate

*  MFC for Nitrogen

*  H2 Gas detectors

*  Additional temperature controller with measurement

*  Additional Temperature measurement

*  Humidity measurement

*  Additional Pressure measurement

*  (Multichannel) Impedance Monitor

*  External potentiostat/galvanostat

*  Automatic back pressure regulator instead of manual type back pressure regulator

*  Zero voltage booster

*  Faradaic oven

*  Fuel cell stack jig

*  Stack cooling system

*  Stack multichannel temperature monitor (each cell)

*  Stack multichannel voltage monitor (each cell)

*  Differential pressure gauge

*  Moisture trap (auto/ manual)

*  DI water filter module

*  Conductivity measurement with DI water chamber

*  Fuel cell hardware fixture

*  L type housing (with/without fuel chamber)

 

*         Hardware Specification

 

*  MFC (Mass Flow Controller), 2 ea

*   Max. Flow rate : H2 2slpm, Air 5slpm

*   Accuracy : 1% Full scale

*  100W Single Cell Application

*   Anode gas flow rate : 0.02 2 slpm

*   Cathode gas flow rate : 0.1 5 slpm

*  Humidifier, 2 ea

*   membrane type(Nafion membrane tube length: 1meter)

*   Automatic Water feeding system(PC control)

*  Built-in Precision Electronic Load

*   4 ranges(voltage & current)

*   Max 200Watt (Max. 100Amp, Max 50V)

*   Standard: 10V(1,2,5,10V) 50A(5,10,50,100Amp)
          Other ranges available.

*   Control: CC, CV, CP,STEPI,STEPV,STEPP.

*  Humidify or dry gas selectable

*   auto 3 way valve

*  Back Pressure regulator, 2ea

*   0-50psi

*  Precision Rotary Piston Pump

*   0.1 45ml/min

*   for DM application

*  Temperature Control System

*   Control & measurement

*   7 points

*   Humidifier, Gas line(inside, outside), Cell Temp

*  Temperature measurement only

*   pipe line inside temperature : 2 points

*  Pressure measurement

*   2 points

*  Compact size

     - 430W x 504H x 430D (mm)

 


 

*         Block Diagram

SMART2TM

Click to enlarge